We compared the peripheral refractive measurements of a recently proposed laser-scanning instrument with an established peripheral refractor. Two-dimensional refractive maps were obtained using both instruments for 18 young subjects with differing values of central refraction. The comparison shows a strong correlation between devices in the overlapping measurement area, with the new device extending the range of the explored retinal area to a 100-degree-diameter circular patch, compared to the 60°x35° rectangular area of the older peripheral refractor. Larger refractive maps exhibit trends that cannot be easily predicted from narrower scans. These results demonstrate that the new instrument can be a useful tool for assessing wide-angle peripheral optical data in the human eye.
Search
Categories
Archives
- December 2024 (4)
- October 2024 (5)
- September 2024 (1)
- July 2024 (8)
- June 2024 (1)
- March 2024 (1)
- October 2023 (2)
- July 2023 (5)
- June 2023 (1)
- July 2022 (10)
- May 2022 (1)
- July 2021 (5)
- March 2021 (2)
- October 2020 (36)
- September 2020 (122)
- August 2020 (10)
- July 2020 (38)
- April 2020 (1)
- April 2018 (1)
- September 2017 (2)
Ultra-wide-angle peripheral refraction using a laser-scanning instrument
- Voptica
- Other Publications
Journal:
Biomedical Optics Express
Year:
2024
Authors:
Santiago Sager, Arturo Vicente-Jaen, Zhenghua Lin, Pedro M. Prieto, Zhikuang Yang, Weizhong Lan, and Pablo Artal