Comparison of the Adaptive Optics Vision Analyzer and the KR-1W for measuring ocular wave aberrations

Journal:
Clinical and Experimental Optometry
Year:
2017
Link:
Authors:
Carles Otero, Meritxell Vilaseca, Montserrat Arjona, Joan A. Martínez-Roda, Jaume Pujol

Background: The aim was to assess the agreement in the measurement of ocular aberrations between a new Adaptive Optics Vision Analyzer (AOVA, Voptica, Murcia, Spain) and a commercial aberrometer (KR-1W, Topcon, Tokyo, Japan), both based on the Hartmann–Shack technique.
Methods: One experienced examiner measured 29 healthy right eyes nine consecutive times with the two instruments. The individual Zernike coefficients and the rootmean square (RMS) of each order from the second to the fifth order, the higher-order RMS (RMSHOA), the total RMS (RMSTOT) and the values of the spherical equivalent (M) and Jackson cross-cylinder (J0 and J45) were compared. All aberrations were computed for a 4.0mm pupil diameter.
Results: Bland and Altman analysis showed good agreement between instruments and most of the parameters showed no statistically significant differences. Although the largest mean differences were obtained for the defocus coefficient C(2,0) and the spherical equivalent (M) with a mean difference (and standard deviation) of 0.190 ± 0.099 μm and -0.150 ± 0.188 D, respectively, they were clinically acceptable and significant correlations were found between the AOVA and KR-1W for the major refractive components such as spherical equivalent (r = 0.995, p<0.001), J0 (r = 0.964, p<0.001), J45 (r = 0.901, p<0.001) and C(4,0) (r = 0.575, p = 0.001).
Conclusions: The results suggest good agreement between instruments. Accommodation and misalignment of themeasurementsmay play a role in some of the statistically significant differences that were obtained, specifically for defocus C(2,0), vertical coma C(3,1) and spherical aberration C(4,0) coefficients; however, these differences were clinically irrelevant.