Purpose: The aim was to evaluate the depth of focus and visual quality in patients who received two refractive IOLs, EDF ArtIOLs® 40 and FVR ArtIOLs® 70 (ArtIOLs, Voptica SL, Murcia, Spain). These lenses generate low (ArtIOLs® 40) and moderate (ArtIOLs® 70) negative spherical aberration, enhancing depth of focus, while the inverted meniscus design improves peripheral optical performance.
Setting: Oftalvist Murcia, Spain
Methods: 83 patients were implanted bilaterally with ArtIOLs® 40 and ArtIOLs® 70, aiming to provide full range of vision. One month best corrected monocular and binocular defocus curve was measured from +1.00 to -3.00 in 0.50 D steps. Monocular and Binocular UCVA were measured at far, 66 and 40 cm. Measurements were taken using SLOAN charts under photopic conditions. Additionally, 64 patients completed the Patient Reported Spectacle Independence Questionnaire (PRSIQ).
Results: For ArtIOLs® 40, monocular BCVA at far, intermediate and near were -0.05, 0.12 and 0.35, respectively, while for ArtIOLs® 70 were -0.05, 0.02 and 0.20 logMAR. Binocular BCVA were -0.06, 0.00 and 0.17 at far, intermediate and near distance. Monocular UCVA for far was -0.02 and 0.03 logMAR, 0.07 and -0.01 at 66 cm and 0.30 and 0.10 logMAR at 40 cm for ArtIOLs® 40 and ArtIOLs® 70, respectively. Binocular UCVA was 0.15 logMAR or better at all distances. All patients reported to be comfortable at all distances all time or most of the time without glasses.
Conclusions: ArtIOLs® 40 and ArtIOLs® 70 fulfil the requirements of the new ISO 11979-7 for Extended Deph of Focus IOLs and Full Visal Range IOLs, respectively. Furthermore, the patients implanted binocularly with ArtIOLs® 40 and ArtIOLs® 70 achieved excellent vision at all distances. High percentage of spectacle independence was achieved.
Buscar
Categorías
Archivos
- diciembre 2025 (2)
- octubre 2025 (7)
- junio 2025 (2)
- mayo 2025 (1)
- marzo 2025 (4)
- diciembre 2024 (4)
- octubre 2024 (5)
- septiembre 2024 (1)
- julio 2024 (8)
- junio 2024 (1)
- mayo 2024 (1)
- marzo 2024 (1)
- octubre 2023 (2)
- julio 2023 (5)
- junio 2023 (1)
- julio 2022 (10)
- mayo 2022 (1)
- marzo 2022 (1)
- julio 2021 (5)
- marzo 2021 (2)
- octubre 2020 (36)
- septiembre 2020 (122)
- agosto 2020 (10)
- julio 2020 (38)
- abril 2020 (1)
- diciembre 2019 (1)
- abril 2018 (1)
- septiembre 2017 (2)
Extended Depth Of Focus And Visual Quality In Full Visual Range Inverted Meniscus IOLs
- Voptica
- Publicaciones de ArtIOLs, Resúmenes ESCRS
Journal:
ESCRS 2025
Year:
2025
Link:
Authors:
Pablo Artal, Jose María Marín, Encarna Alcón, Consuelo Robles, Lucía Hervella, Elsa Albero, Eloy Villegas
PDF: